The HI98713 is a high accuracy portable turbidity meter. The meter is supplied complete with AMCO-AEPA-1 primary turbidity standards used for calibration and performance verification. The HI98713 meets and exceeds the requirements of the ISO 7027 Method for turbidimetric measurements.
The HI98713 Precision ISO Turbidity Portable Meter is specially designed for water quality measurements, providing reliable and accurate readings, even within low turbidity ranges. The instrument is based on a state-of-the-art optical system which guarantees accurate results, assures long term stability, and minimizes stray light and color interferences. Periodic calibration with the supplied standards compensates for any variations in intensity of the LED light source. The 25 mm round cuvettes composed of special optical glass guarantee the repeatability of turbidity measurements.
Features at-a-glance
Ratio Measurement Mode – The HI98713 measures turbidity using the ratio method with a 90° and 180° light detector for more accurate measurements. Turbidity measurements can be made in the 0.00 to 1000 FNU range.
Multiple reading modes – Normal measurement, continuous measurement, or signal averaging measurement are reading modes available.
ISO Compliant – The HI98713 meets and exceeds the requirements of ISO 7027 method for turbidity measurements by use of an infrared LED light source.
Calibration – The HI98713 has a powerful calibration function that compensates for variation in light intensity. The calibration can be done using the supplied calibration solutions or user-prepared standards. A two, three, or four-point turbidity calibration can be performed by using the supplied (<0.1, 15, 100 and 750 FNU) standards. Calibration points can be modified if user-prepared standards are used.
AMCO AEPA-1 Primary Turbidity Standard – The AMCO AEPA-1 supplied standards are recognized as a primary standard by the USEPA. These non-toxic standards are made of styrene divinylbenzene polymer spheres that are uniform in size and density. The standards are reusable and stable with a long shelf life.
Fast Tracker™ – For advanced field applications, the HI98713 is equipped with Fast Tracker™ – Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a specific measurement or series of measurements using iButton® tags near sampling points for quick and easy readings. Each iButton® tag contains a computer chip with a unique identification code encased in stainless steel.
GLP Data – The HI98713 features complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. Data includes calibration points, date, and time.
Data Logging – Up to 200 measurements can be stored in the internal memory and recalled at any time.
Data Transfer – For further storage or analysis options, logged data can be downloaded to a Windows® compatible PC using the USB or RS232 port and the HI92000 software.
Backlit Graphic LCD Display – A graphic LCD display provides an easy to understand, user-friendly interface. All messages are in plain text making them easy to read.
Significance of Use
Turbidity is one of the most important parameters used to determine the quality of drinking water. Once considered as a mostly aesthetic characteristic of drinking water, significant evidence exists that controlling turbidity is a competent safeguard against pathogens. In natural water, turbidity measurements are taken to gauge general water quality and its compatibility in applications involving aquatic organisms. The monitoring and treatment or wastewater was once solely based on the control of turbidity. Currently, the measurement of turbidity at the end of the wastewater treatment process is necessary to verify that the values are within regulatory standards.
Turbidity of water is an optical property that causes light to be scattered and absorbed, rather than transmitted. The scattering of light that passes through a liquid is primarily caused by the suspended solids present. The higher the turbidity, the greater the amount of scattered light. Even a very pure liquid will scatter light to a certain degree, as no solution will have zero turbidity.
The ISO standard for turbidity measurement uses infrared wavelength of light that is outside the visible spectrum. The primary benefit of the ISO method over the EPA method is the reduction of color interference. The EPA method uses a tungsten lamp that produces light that contains all of the visible wavelengths of light that we see as white. A solution that is colored will be adsorbed by a complementary wavelength of light which will affect the turbidity reading. Since the ISO method is outside the visible wavelength of light the color in the sample does not interfere with the measurement. The USEPA prefers the tungsten lamp since it provides for better accuracy at low ranges and drinking water should not be colored. The requirements of an optical turbidity meter for the measurement of diffuse radiation used in low range measurements (i.e. drinking water) expressed as formazin nephelometric units (FNU) according to ISO 7027 are:
Measurement of diffuse radiation for water with low turbidity (0 FNU to 40 FNU)
The wavelength of the incident radiation shall be 860 nm
The spectral bandwidth of the incident radiation shall be less than or equal to 60 nm
There shall be no divergence from parallelism of the incident radiation and any convergence shall not exceed 1.5°.
The measuring angle, theta, between the optical axis of the incident radiation and that of the diffused radiation shall be 90° ±2.5°.
The aperture angle should be between 20° and 30° in the water sample.
The HI98713 meets and exceeds the meter criteria specied by ISO 7027.
Specification of ISO Turbiditymeter HI98713 Hanna Instruments :
Specification Name
Detail
sku
HI98713-01
FNU Range
0.00 to 9.99 FNU
10.0 to 99.9 FNU
100 to 1000 FNU
FNU Resolution
0.01 FNU from 0.00 to 9.99 FNU
0.1 FNU from 10.0 to 99.9 FNU
1 FNU from 100 to 1000 FNU
FNU Accuracy
±2% of reading plus 0.1 FNU
Range Selection
Automatic
Turbidity Repeatability
±1% of reading or 0.01 FNU, whichever is greater
Stray Light
< 0.1 FNU
Typical EMC Deviation
±0.05 FNU
Turbidity Light Detector
860 nm infrared LED
IR Detector
Silicon Photocell
Turbidity Method
Adaptation of ISO 7027, ratio method with 90° and 180° detector.
Turbidity Standards
< 0.1, 15, 100 and 750 FNU
Calibration
Two, three or four point calibration
Display
60 x 90 mm LCD with backlight
LOG Memory
200 records
Serial Interface
RS232 and USB 1.1
Environment
0 to 50°C (32 to 122°F); max 95% RH non-condensing
ISO Turbiditymeter HI98713 Hanna Instruments
Download Catalog :
The HI98713 is a high accuracy portable turbidity meter. The meter is supplied complete with AMCO-AEPA-1 primary turbidity standards used for calibration and performance verification. The HI98713 meets and exceeds the requirements of the ISO 7027 Method for turbidimetric measurements.
Description
ISO Turbiditymeter HI98713 Hanna Instruments. Lihat juga Turbiditymeter HI98703.
The HI98713 Precision ISO Turbidity Portable Meter is specially designed for water quality measurements, providing reliable and accurate readings, even within low turbidity ranges. The instrument is based on a state-of-the-art optical system which guarantees accurate results, assures long term stability, and minimizes stray light and color interferences. Periodic calibration with the supplied standards compensates for any variations in intensity of the LED light source. The 25 mm round cuvettes composed of special optical glass guarantee the repeatability of turbidity measurements.
Features at-a-glance
Ratio Measurement Mode – The HI98713 measures turbidity using the ratio method with a 90° and 180° light detector for more accurate measurements. Turbidity measurements can be made in the 0.00 to 1000 FNU range.
Multiple reading modes – Normal measurement, continuous measurement, or signal averaging measurement are reading modes available.
ISO Compliant – The HI98713 meets and exceeds the requirements of ISO 7027 method for turbidity measurements by use of an infrared LED light source.
Calibration – The HI98713 has a powerful calibration function that compensates for variation in light intensity. The calibration can be done using the supplied calibration solutions or user-prepared standards. A two, three, or four-point turbidity calibration can be performed by using the supplied (<0.1, 15, 100 and 750 FNU) standards. Calibration points can be modified if user-prepared standards are used.
AMCO AEPA-1 Primary Turbidity Standard – The AMCO AEPA-1 supplied standards are recognized as a primary standard by the USEPA. These non-toxic standards are made of styrene divinylbenzene polymer spheres that are uniform in size and density. The standards are reusable and stable with a long shelf life.
Fast Tracker™ – For advanced field applications, the HI98713 is equipped with Fast Tracker™ – Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a specific measurement or series of measurements using iButton® tags near sampling points for quick and easy readings. Each iButton® tag contains a computer chip with a unique identification code encased in stainless steel.
GLP Data – The HI98713 features complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. Data includes calibration points, date, and time.
Data Logging – Up to 200 measurements can be stored in the internal memory and recalled at any time.
Data Transfer – For further storage or analysis options, logged data can be downloaded to a Windows® compatible PC using the USB or RS232 port and the HI92000 software.
Backlit Graphic LCD Display – A graphic LCD display provides an easy to understand, user-friendly interface. All messages are in plain text making them easy to read.
Significance of Use
Turbidity is one of the most important parameters used to determine the quality of drinking water. Once considered as a mostly aesthetic characteristic of drinking water, significant evidence exists that controlling turbidity is a competent safeguard against pathogens. In natural water, turbidity measurements are taken to gauge general water quality and its compatibility in applications involving aquatic organisms. The monitoring and treatment or wastewater was once solely based on the control of turbidity. Currently, the measurement of turbidity at the end of the wastewater treatment process is necessary to verify that the values are within regulatory standards.
Turbidity of water is an optical property that causes light to be scattered and absorbed, rather than transmitted. The scattering of light that passes through a liquid is primarily caused by the suspended solids present. The higher the turbidity, the greater the amount of scattered light. Even a very pure liquid will scatter light to a certain degree, as no solution will have zero turbidity.
The ISO standard for turbidity measurement uses infrared wavelength of light that is outside the visible spectrum. The primary benefit of the ISO method over the EPA method is the reduction of color interference. The EPA method uses a tungsten lamp that produces light that contains all of the visible wavelengths of light that we see as white. A solution that is colored will be adsorbed by a complementary wavelength of light which will affect the turbidity reading. Since the ISO method is outside the visible wavelength of light the color in the sample does not interfere with the measurement. The USEPA prefers the tungsten lamp since it provides for better accuracy at low ranges and drinking water should not be colored. The requirements of an optical turbidity meter for the measurement of diffuse radiation used in low range measurements (i.e. drinking water) expressed as formazin nephelometric units (FNU) according to ISO 7027 are:
Measurement of diffuse radiation for water with low turbidity (0 FNU to 40 FNU)
The HI98713 meets and exceeds the meter criteria specied by ISO 7027.
Specification of ISO Turbiditymeter HI98713 Hanna Instruments :
10.0 to 99.9 FNU
100 to 1000 FNU
0.1 FNU from 10.0 to 99.9 FNU
1 FNU from 100 to 1000 FNU
Related products
Turbiditymeter HI93703-11 Hanna Instruments
Haze Turbiditymeter for Beer HI847492 HANNA INSTRUMENTS