HI83305 is a compact, multiparameter photometer intended for boiling and cooling tower applications for use in the lab or in the field. The meter is one of the most advanced photometers available with an innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. This meter has 30 different programmed methods measuring 18 key water quality parameters and also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves. Parameters specific to boiling and cooling towers include oxygen scavengers and silica which are important for maintaining equipment. To save valuable laboratory benchtop space, the HI83305 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.
Application Notes
The HI83305 benchtop photometer measures 18 different key water quality parameters using 30 different methods. This photometer features an innovative optical system that uses LEDs, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.
Specially designed for use with boilers and cooling towers, the HI83305 is a comprehensive way to maintain precise water conditions in systems. Problems such as corrosion, deposition, and microbial growth can occur if these key parameters, such as oxygen scavengers and silica, arent maintained. Oxygen scavengers are added to remove residual dissolved oxygen in boiler feed water that can cause corrosion in a steam generating plant. It is important that levels of oxygen scavengers be routinely checked to prevent against corrosion and ensure that equipment is working efficiently. Boiler water maintenance is necessary to prevent or control deposit formation as seen with silica. Silica contamination can reduce system efficiency and increase maintenance of equipment due to scaling.
A digital pH electrode input allows the user to measure pH by a traditional glass electrode. The digital pH electrode has a built-in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built-in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.
The HI83305 offers an absorbance measuring mode that allows for CAL Check TM standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the five wavelengths of light (420 nm, 466 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for educators to teach the concept of absorbance by using the Beer-Lambert Law.
Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.
Specification of HI83305 Boiler and Cooling Tower Photometer
Specification
Detail
sku
HI83305-02
Wavelength Accuracy
±1 nm
Light Source
5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters
Spectral bandwidth
8 nm
Input Channels
1 pH electrode input and 5 photometer wavelengths
pH Electrode
digital pH electrode (not included)
Logging Type
log on demand with user name and sample ID optional input
Logging Memory
1000 readings
Connectivity
USB-A host for flash drive micro-USB-B for power and computer connectivity
GLP
calibration data for connected pH electrode
Display
128 x 64 pixel LCD with backlight
Battery Type / Life
3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement
Power Supply
5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Environment
0 to 50.0oC (32 to 122.0oF) 0 to 95% RH, non-condensing
Dimensions
206 x 177 x 97 mm (8.1 x 7.0 x 3.8)
Weight
1.0 kg (2.2 lbs.)
Photometer/Colorimeter Light Detector
silicon photodetector
Cuvette Type
round, 24.6 mm
Number of Methods
128 max.
Ordering Information
Supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual.
HI83305 Boiler and Cooling Tower Photometer Hanna Instruments
Download Catalog :
Description
HI83305 Boiler and Coolling Tower Photometer Hanna Instruments.
Kami PT. Alfa Omega Indolab merupakan distributor untuk alat-alat Laboratorium. Kami jual Photometer HI83305 Photometer untuk Boiler dan Cooling Tower serta berbagai macam pH Meter Hanna Instrument dengan dukungan purna jual prima sehingga memberikan kenyaman kepada user. Melayani pengiriman ke seluruh kota di Indonesia. Temukan produk-produk kami yang lain untuk melengkapi kebutuhan Laboratorium Anda. Berikut beberapa alat ukur penting untuk uji kualitas air : pH Meter Hanna Instruments, TDS Meter, DO Meter, Photometer, COD Meter, Turbidity Meter, Conductivity Meter, Salinity Meter, ORP Meter, ISE Meter, Refractometer, Thermometer, Chemical Test Kit, dll. Hanna Instrument fokus memproduksi alat ukur kualitas air, sehingga sangat menjaga mutunya.
General Specification :
HI83305 is a compact, multiparameter photometer intended for boiling and cooling tower applications for use in the lab or in the field. The meter is one of the most advanced photometers available with an innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. This meter has 30 different programmed methods measuring 18 key water quality parameters and also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves. Parameters specific to boiling and cooling towers include oxygen scavengers and silica which are important for maintaining equipment. To save valuable laboratory benchtop space, the HI83305 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.
Application Notes
The HI83305 benchtop photometer measures 18 different key water quality parameters using 30 different methods. This photometer features an innovative optical system that uses LEDs, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.
Specially designed for use with boilers and cooling towers, the HI83305 is a comprehensive way to maintain precise water conditions in systems. Problems such as corrosion, deposition, and microbial growth can occur if these key parameters, such as oxygen scavengers and silica, arent maintained. Oxygen scavengers are added to remove residual dissolved oxygen in boiler feed water that can cause corrosion in a steam generating plant. It is important that levels of oxygen scavengers be routinely checked to prevent against corrosion and ensure that equipment is working efficiently. Boiler water maintenance is necessary to prevent or control deposit formation as seen with silica. Silica contamination can reduce system efficiency and increase maintenance of equipment due to scaling.
A digital pH electrode input allows the user to measure pH by a traditional glass electrode. The digital pH electrode has a built-in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built-in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.
The HI83305 offers an absorbance measuring mode that allows for CAL Check TM standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the five wavelengths of light (420 nm, 466 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for educators to teach the concept of absorbance by using the Beer-Lambert Law.
Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.
Specification of HI83305 Boiler and Cooling Tower Photometer
Related products
HI97115 Marine Multiparameter Photometer
HI97749 Chromium Low Range Photometer Hanna Instruments
HI83306 Environmental Analysis Photometer Hanna Instruments